Parr-instrument A3040HC3 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Parr-instrument A3040HC3. Parr Instrument A3040HC3 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 4
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
www.parrinst.comParr Instrument Company
A3040HC3 Compact Magnetic Drive
Operating Instruction Manual
655M
Description
The A3040HC3 compact magnetic stirrer consists of
a lower rotor assembly with an inner set of magnets
and an outer coupling assembly with a second set of
magnets. This assembly is intended for low viscos-
ity applications and has a torque rating of 2.5 in-lbs
(28.2 Ncm).
The A3040HC3 magnetic stirrer can be used with
both the standard stirrer shaft and impeller(s) or
with a gas entrainment stirrer arrangement. It is
designed to stir at speeds up to approximately 1700
rpm. Maximum stirrer speeds are dependent upon
the viscosity of the vessel contents and the resultant
load on the motor.
This magnetic stirrer requires minimal maintenance;
however, it is important to protect these magnetic
stirrers from the heat in the reactor vessel. To pre-
vent overheating each magnetic drive includes a
cooling sleeve. The cooling sleeve is held in place
on the housing with a snap ring and seals against
the housing with two O-rings. A steady fl ow of cold
water through this sleeve during all vessel operation
at temperatures above 80 ºC is required. Excessive
heating without use of the cooling water will also
damage the O-rings in the cooling sleeve.
Shaft Removal and Installation
To remove the stirrer shaft hold the A3077HC outer
housing of the compact magnetic drive assembly,
and unscrew the stirrer shaft in a counter-clockwise
rotation. To install a replacement or alternate shaft,
hold the outer housing of the compact magnetic
drive fi rmly and screw in the new shaft.
*Note: It is important to hold the housing
securely when installing any of the stirrer
shafts so that the threads are fully engaged.
Magnetic Drive Maintenance
The thrust washers and bushings in this magnetic
stirrer are self lubricating and do not require service
during normal operation.
It is advisable to periodically inspect the internal
portion of this drive to monitor the wear on both the
thrust washers and bushings as well as any debris
build-up that may occur from solvents in the reac-
tor system. This can be done without removing the
entire drive from the vessel head.
To inspect the internal rotor area:
Loosen the (4) set screws located around the outer
housing. Lift the A3077HC outer housing assem-
bly off of the magnetic drive. Unscrew the end cap
(A3076HC) out of the magnetic drive. This will expose
the 1137HCHA PTFE O-ring as well as the lower rotor,
thrust washers, and bushings.
Check for any evidence of leakage past the O-ring
seal. If vapors from the reaction have produced
solid deposits around the rotor housing these will
need to be removed to maintain free movement of
the inner rotor assembly.
To remove the A3035HC internal rotor assembly,
push upward using the internal stirrer shaft. Hold
the internal rotor and unscrew the stirrer shaft from
the rotor.
The two thrust washers are positioned on the top
and bottom sides of the internal rotor. The lower two
bushings are located in the 2721HC3 housing. The
upper (3rd) bushing is located in the bottom of the
end cap. These components require periodic replace-
ment, typically once or twice a year depending upon
usage. Over time the bushings will wear leaving
small fl akes of material in the housing. It will be
evident with visual inspection of these thrust wash-
ers and bushings when replacement is required. The
1132HC upper and lower thrust washers, as well as
the three 1133HC bushings, are all manufactured of
carbon-fi lled PTFE and are easily replaced.
Vue de la page 0
1 2 3 4

Résumé du contenu

Page 1 - Operating Instruction Manual

www.parrinst.comParr Instrument CompanyA3040HC3 Compact Magnetic DriveOperating Instruction Manual655MDescriptionThe A3040HC3 compact magnetic stirrer

Page 2 - Parr Instrument Company

A3040HC3 Compact Magnetic DriveParr Instrument Company2Bushing RemovalRemove the A3035HC inner rotor assembly from the 2721HC3 housing, the upper thru

Page 3 - Cooling Sleeve Assembly

A3040HC3 Compact Magnetic Drivewww.parrinst.com3A3040HC3 Magnetic Drive Parts ListsItem DescriptionA3076HC* Bearing End Cap AssemblyA3077HC Outer Mag

Commentaires sur ces manuels