The 6100 Calorimeter will hold data for 1000 tests in its memory. These tests may be
pre weights, preliminary or final reports for either standardization or determination
runs. Once the memory of the controller is filled, the controller will not start a new
analysis until the user clears some of the memory.
Clearing Memory
The FILE MANAGEMENT key on the main menu
leads to the file management sub-menu. The RUN
DATA FILE MANAGER key leads to a listing of the
files.
Single files can be deleted by highlighting the
file and pressing the DELETE key. The controller
will then ask the user to confirm that this file is
to be deleted.
A series of files can be deleted by selecting the
first file in the series and then the last file in
the series using the EXTEND SEL key and then
pressing the DELETE key.
Removable SD Memory Cards
The controller of the 6100 Calorimeter can accept SD
memory cards. These cards can be used to:
Copy test file data for transfer to a computer.
Copy user settings for back up.
Reload user settings to the controller.
Restore or update the controllers operating
system.
SD memory cards are inserted into one of the slots
on the back of the control section of the calorimeter.
Keys are provided on the FILE MANAGEMENT sub-
menu to initiate each of the above three actions with
the exception of restoring or updating the control-
lers operating system.