HTML Installation
www.parrinst.com
5
Locate the free yellow wire #5 from the kit and 8.
attach one end to pin 10 on the 2082E meter. The
other end attaches to the 1119E solid state relay
(position 3).
SSR Position 3
Pin Outs:
2082E Color: Attaches to:
Pin 1 Black Terminal Block 5
Pin 2 White Terminal Block 2
Pin 3
Pin 4 White T/C jack +
Pin 5
Pin 6 Red T/C jack -
Pin 7 Blue Terminal Block 7
Pin 8 Orange Terminal Block 6
Pin 9 Red 2083E Primary Temp
Meter
Pin 10 Yellow 1119E SSR (3)
Pin 11 White Terminal Block 4
Pin 12 Black Terminal Block 3
* White (Type – J), Yellow (Type – K), Blue (Type – T),
Black (RTD)
** Red (Type – J, Type – K, Type – T), White (RTD)
Final Steps:
Close the controller and replace the two screws on
the top plate. Plug the 4848 controller back in, and
turn it on.
It is useful to check that the settings on the display
are set correctly. Check these against the defaults
listed in the back of these instructions.