Cal 9500P
Installation Instructions
2 Parr Instrument Company
10. Connect the wires of the Cal 9500P
temperature controller as follows: (See Figure 4)
Wire:
Cal Pin #
Position:
Function
Black 7 Power
White 8 Power
Orange 24 High limit
Blue 23 High limit
Brown 21 SVM
Yellow 22 SVM
Red 19 SSR
Purple 20 SSR
T/C White
(Type J)
1 T/C
T/C Red
(Type J)
2 T/C
11. Remove the four wires connected to the
solid state relay on the back panel of the 4840
controller. (See Figure 5)
12. Remove the solid state relay from the back
of the 4840 controller and resistor if present.
Discard the old solid-state relay and resistor.
13. Install the provided solid-state relay (part #:
1119E) on the back panel of the 4840 controller.
(See Figure 6)
14. Attach the short black wire to position 1 of
the relay.
15. Attach the long black wire to position 2 of the
relay.
16. Attach the red wire to position 3 of the relay.
17. Attach the purple (or blue) wire to position 4
of the relay.
Back of Cal 9500P with wiring
Figure 4
Solid state relay for use with Watlow
(to be replaced)
Figure 5
Commentaires sur ces manuels