Parr-instrument 1104 Manuel d'utilisateur Page 7

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 12
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 6
1104 Oxygen Combustion Vessel
www.parrinst.com
7
threaded on instead of pushed on to the inlet valve.
The correct assembly to use when lling the 1104
from a calorimeter is P/N A19A15. Close the outlet
valve on the bomb head. Thread the end of the ll
connection on to the inlet valve of the 1104. Press
O2 Fill” on the calorimeter to ll the vessel.
Filling the Bomb Manually
To ll the vessel directly the 1823 Oxygen Fill Con-
nection must be used. Close the outlet valve on the
bomb head. Open the oxygen tank valve not more
than a quarter turn. Open the control valve on the
ll connection slowly and watch the bomb pres-
sure rise to the desired lling pressure. Close the
control valve on the ll connection. The inlet check
valve will close automatically when the residual
pressure in the lling hose is released; leaving the
bomb lled to the highest pressure indicated on
the gage. Release the residual pressure in the ll-
ing hose by pushing downward on the toggle relief
lever attached to the ll connection. The gage should
now drop to zero. If the pressure drops slowly and
a large amount of gas escapes when the pressure
relief valve is opened, the check valve in the bomb
head is not operating properly. This trouble will have
to be corrected before the bomb can be used.
If too much oxygen should accidentally be intro-
duced into the bomb, do not proceed with the
combustion. Detach the lling connection, exhaust
the bomb, remove the head and reweigh the sample
before repeating the lling operation.
When using the bomb in a calorimeter, insert the
lifting handle into the two holes in the side of the
screw cap and lower the bomb partially into the
calorimeter water bucket. Note that the bucket
should be lled with 1850 mL of water. Press the
banana plugs on the two ignition wires rmly into
the terminal sockets on the bomb head before the
head is immersed in the water. After connecting the
wires lower the bomb into the bucket. Remove the
lifting handle and shake off any drops of water back
into the bucket. The calorimeter should be standard-
ized using an amount of Benzoic Acid that releases
approximately the same amount of energy as the
sample to be tested. An approximate EE value for
the 1104 in the 6000 Series instruments is 2500
cal/°C.
When using the bomb alone, it must be connected
to an ignition unit (P/N 2901). Always place the 1104
Oxygen Combustion Vessel behind a heavy shield
or barricade and completely submerge the bomb in
cold running water while ring. If the sample devel-
ops an unusually large amount of heat, it is recom-
mended that a cold stream of water be run against
the bomb to facilitate cooling.
Note: Do not fire the bomb at any point if there
is gas leakage.
The bomb must remain in the water bath until all
parts of the head and screw cap come to tempera-
ture equilibrium. Then remove the bomb from the
bath, wipe it with a clean towel and release the
residual gases through the needle valve before
unscrewing the compression screws and removing
the cap. If necessary, a spanner wrench (P/N 219A) is
provided to remove the screw cap of the vessel.
Maintenance
The chromium plated, steel alloy screw cap on the
bomb requires special care to keep the threads
from rusting and to prevent seizure. Always dry the
cap thoroughly after it has been used and store the
bomb with the screw cap removed from the cylin-
der. Keep a light coating of anti-seize lubricant (P/N
424HC2) on the threads of the cap but do not use
this or any lubricant on any other parts of the bomb.
Basic maintenance of the bomb including replacing
the O-ring, insulators, gaskets and spacers is recom-
mended every 500 rings. If the bomb is used for
samples that contain a high amount of corrosive
materials, this maintenance should be completed
after every 250 rings.
Parr provides Oxygen Bomb repair services for the
1104 (P/N REPAIR.O7). A factory test is recommend-
ed after 5000 rings. This includes replacing all of
the seals, checking the dimensions of the cylinder,
screw cap and head, hydrostatically testing the seals
and proof ring the vessel using benzoic acid.
When returning a bomb to the factory, ship it to:
Parr Instrument Company
Attn: Repair Department
211 53rd Street
Moline, IL 61265-1720
Include a Purchase Order to cover the cost of the
repair, a person to contact (complete with a phone
number) and return shipping information. Individual
repair parts can be ordered from any Parr dealer or
directly from Parr.
Vue de la page 6
1 2 3 4 5 6 7 8 9 10 11 12

Commentaires sur ces manuels

Pas de commentaire