6725 Semi-micro Calorimeter Instruction Manual
6-6
Note:
The values entered into the test report appear as entered in the report. Values for
e
1
, e
2
and e
3
are calculated and used as energy corrections in accordance with the
formulas and settings given above. The formulas used above to arrive at e
1
or e
2
are not the same as the formulas used for e
1
and e
2
which appear in most ASTM
bomb calorimetric procedures. However, the sum of e
1
and e
2
, above, is equal to
the sum of the ASTM treatment of e
1
and e
2.